Agilent 320-MS Triple Quadrupole Mass Spectrometer Application Note 01992

This document introduces a method for quantifying multi-class pharmaceuticals in water using simultaneous positive and negative ESI on the Varian 320-MS triple quadrupole mass spectrometer


Agilent 320-MS Triple Quadrupole Mass Spectrometer Application Note 01992 - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
91 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
320-MS
MD5 checksum
4EADB8BD3B241B21808C725041F62B2A

About this manual

This document details a method for quantifying twelve multi-class pharmaceuticals in water using the Agilent 320-MS triple quadrupole mass spectrometer. The study utilizes simultaneous positive and negative Electrospray Ionization (ESI) in Multiple Reaction Monitoring (MRM) mode to analyze a broad range of compounds, including antibiotics and endocrine disruptors, in treated wastewater and surface water.

The application outlines specific Liquid Chromatography (LC) and Mass Spectrometry (MS) parameters, including column conditions, solvent gradients, and MRM transitions for analytes such as Amoxicillin, Acetaminophen, and Ibuprofen. Results indicate that the method achieves excellent separation in less than 10 minutes with high linearity (r² > 0.995) and acceptable repeatability (%RSD < 13%) for the tested pharmaceuticals.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • LC Conditions
  • MS Parameters
  • Results and Discussion
  • Calibration Data
  • Conclusion

Specifications

Drying Gas25 psi at 310 °C
Nebulizing Gas55 psi
Needle Voltage5000 V
Shield Voltage600 V
Calibration Range1-50 ng/mL
ColumnPursuit XRs C18, 3 um, 50 x 2.0 mm
Run Timeless than 10 min

Frequently asked questions

What is the calibration range for the pharmaceuticals tested?

The calibration results are presented over a range of 1-50 ng/mL.

What is the typical run time for the analysis?

The method achieves a run time of less than 10 minutes.