Agilent Star 9080 EC Detector Catecholamine Analysis Application Note

The document introduces the application of Star 9080 Amperometric Electrochemical Detector . The detector is used to analyze the standards of 30 nmol/L noradrenaline (1), adrenaline (2) and dopamine (3).


Agilent Star 9080 EC Detector Catecholamine Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
119 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
B08C63002543C2BD0BCE01131C1A7226

About this manual

This document details the analysis of catecholamines (noradrenaline, adrenaline, and dopamine) using the Varian Star 9080 Amperometric Electrochemical Detector. It specifies the standard separation conditions, including the use of an ODS column, a mobile phase containing phosphoric acid, citric acid, EDTA, and octane sulphonic acid, and a flow rate of 1.0 mL/min.

What's inside

  • Standard Separation
  • Calculation of S/N Ratio and Detection Limit
  • Neurotransmitters in rat tissue

Specifications

Detector TypeAmperometric Electrochemical Detector
Flow Rate1.0 mL/min
Column Dimensions150 x 4.6 mm
Injection Volume20 uL
Flowcell Electrode2.74 mm Glassy Carbon
Reference ElectrodeAg/AgCl filled with saturated KCI/AgCl
E-cell Potential800 mV
I-cell Current3 nA

Frequently asked questions

What is the typical S/N ratio for norepinephrine measured with a VT-03 glassy carbon flowcell?

A typical S/N ratio for norepinephrine measured with a VT-03 glassy carbon flowcell is 53000.

What is the concentration detection limit based on three times the noise?

The concentration detection limit based on three times the noise is 56 pmol/L.

What is the minimum detectable amount for the described analysis?

The minimum detectable amount is 1.12 fmol.