Varian PolyView Multicomponent Analysis Application Note 15

This application note describes a method for rapid separation and quantitation of a mixture of 10 non-steroidal anti-inflammatory drugs (NSAIDs) using fast high-performance liquid chromatography (HPLC) and diode array detector (DAD). Separations are performed in less than 5 minutes using a 3-cm long column. To enhance resolution, two wavelengths, 254 and 360 nm, are used. Quantitation is performed by integrating the peak areas of the absorbance peaks at each wavelength. This method can be used to rapidly identify and quantify the NSAIDs present in a mixture.


Varian PolyView Multicomponent Analysis Application Note 15 - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
159 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
B0C88B6A74FC5898FDAE297A0170DBD0

About this manual

This document details a method for analyzing mixtures of 10 non-steroidal anti-inflammatory drugs (NSAIDs) using fast HPLC with diode array detection. The primary challenge addressed is the co-elution of components, specifically ibuprofen, oxaprofen, and indomethacin, which overlap in standard chromatographic runs.

The text explains how Varian's PolyView multicomponent analysis (MCA) software utilizes spectral data to identify and accurately quantitate these co-eluting compounds. By analyzing the differences in individual spectra within the fused peaks, the software can resolve components that would otherwise be indistinguishable at a single wavelength, allowing for accurate quantitation without requiring longer, slower columns.

What's inside

  • Introduction
  • Experimental
  • Results and Discussion
  • Analysis of Standard Mixture A
  • Analysis of Standard Mixture B
  • Conclusions

Frequently asked questions

How many coeluting compounds can the PolyView MCA software identify and quantitate?

The software can identify and quantitate up to 6 coeluting compounds.

What is the maximum analysis time for the NSAID mixtures described in this note?

The mixtures of 10 non-steroidal anti-inflammatory drugs were analyzed in less than 4 minutes.

What is the wavelength range used for the multicomponent spectrum analysis of fused peaks 7 and 8?

The analysis was performed over the 220-325 nm wavelength range.