Varian 9200 Prospekt Application Note: Analysis of Bentazone and Phenoxyacid Herbicides

This article presents the experimental results of analyzing phenoxyacid herbicides and bentazone in water samples using automated on-line solid-phase extraction. The method can be used for frequent monitoring of contaminants in the river Rhine and drinking water samples.


Varian 9200 Prospekt Application Note: Analysis of Bentazone and Phenoxyacid Herbicides - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
173 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
8537E29EC5BA21B23994E7CE5773D3EE

About this manual

This document details an automated method developed for monitoring Bentazone and phenoxyacid herbicides in drinking water and river samples, such as those from the Rhine. The method utilizes an automated on-line solid-phase extraction system controlled by the Varian 9200 Prospekt Controller.

What's inside

  • Introduction
  • Instrumentation
  • Chromatographic Conditions
  • Sample Preparation
  • Results

Specifications

Analytical ColumnPRP-1, 150x 4.1 mm, 10 um
Detection Wavelength230 nm
Sample Volume (Load)25 - 50 ml

Frequently asked questions

Which controller is used for automated on-line solid-phase extraction?

The automated on-line solid-phase extraction was performed by the 9200 Prospekt Controller.

What is the detection wavelength for the herbicides?

Detection was performed at a variable wavelength of 230 nm.

What type of cartridge is used for sample preparation?

A PLRP-S solid-phase extraction cartridge with 15-25 um particle size is used.