Agilent CRA-90 Carbon Rod Atomizer Application Note

This application note describes how to use the CRA-90 carbon rod atomizer to determine priority pollutant metals.


Agilent CRA-90 Carbon Rod Atomizer Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
270 KB
Published
12 June, 2012
Updated
01 October, 2023
MD5 checksum
B49F7F35BA0E391EA366CC300E214704

About this manual

The document addresses the analysis of trace metal concentrations in environmental samples, such as drinking water and industrial effluents, which often require the high sensitivity of graphite furnace methods over flame atomic absorption. It focuses on the determination of twelve priority pollutant metals: Silver (Ag), Arsenic (As), Beryllium (Be), Cadmium (Cd), Chromium (Cr), Copper (Cu), Nickel (Ni), Lead (Pb), Antimony (Sb), Selenium (Se), Thallium (Tl), and Zinc (Zn). Mercury (Hg) is noted as requiring a separate cold vapor technique.

The text provides specific operating parameters, including temperature programs, matrix modifiers, and gas settings, for analyzing these metals in various matrices like aqueous standards, acetic acid, drinking water, and high-salt effluent. It presents data on sensitivity, detection limits, precision, and recovery rates, along with discussions on interference correction and calibration strategies.

What's inside

  • Introduction
  • Silver Determination
  • Arsenic Determination
  • Beryllium Determination
  • Sample Matrices
  • Interferences and Corrections
  • Calibration Curves
  • Detection Limits

Specifications

Sensitivity (Silver)0.8 x 10^-12 g
Detection Limit (Silver)0.3 x 10^-12 g
Sensitivity (Arsenic)7.0 x 10^-12 g
Detection Limit (Arsenic)6.5 x 10^-12 g
Sensitivity (Beryllium)0.6 x 10^-12 g
Detection Limit (Beryllium)0.9 x 10^-12 g

Frequently asked questions

Which metals are covered in this application note?

The note investigates twelve priority pollutant metals: Ag, As, Be, Cd, Cr, Cu, Ni, Pb, Sb, Se, Tl, and Zn.

What is the approved method for Mercury (Hg)?

The only approved method for Mercury is the cold vapor technique.

What matrix modifier is used for Arsenic in effluent samples?

A concentration of 5000 pg/mL Nickel is used to stabilize Arsenic in high salt content effluent samples.