Agilent AA-475 Aluminium Speciation Application Note

This document mainly introduces a chemical speciation method for aluminium in environmental water bodies, which is based on bromide extraction separation and ICP-AES for measurement.


Agilent AA-475 Aluminium Speciation Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
43 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
5D77BA9A1EA552148A3246D33AA6C782

About this manual

This document details a procedure for determining aluminium speciation in natural waters, specifically addressing the ecological dangers of acid rain and the toxicity of aluminium to fish. The method utilizes Graphite Furnace Atomic Absorption Spectroscopy (GFAAS) to isolate toxic chemical forms of aluminium that do not correlate with total aluminium levels.

What's inside

  • Background
  • Speciation Procedure
  • Fast Reactive Aluminium
  • Total Reactive Aluminium
  • Dialysis
  • Graphite Furnace Analysis
  • Analytical Conditions
  • Conclusion

Specifications

InstrumentAgilent AA-475 and GTA-95 graphite furnace and autosampler
Purge gasArgon
ModePeak height
Ash stage temperature1225 C
Atomize stage temperature2700 C

Frequently asked questions

What instrument is used for the analysis?

The analysis is carried out using the Agilent AA-475 and GTA-95 graphite furnace and autosampler.

What is the recommended sample volume for water analysis?

A sample volume of 5 uL is pipetted into a 15 mL disposable polypropylene centrifuge tube.

What is the atomization temperature for aluminium determination?

The atomization temperature is raised to 2700 C to increase sensitivity.