Agilent VGA-76 Vapor Generation Accessory Application Note

This article introduces an automated vapor generation accessory for atomic absorption analysis. The accessory utilizes continuous flow technology to pump and mix samples and liquid reagents. It is suitable for determining low levels of elements such as arsenic, antimony, bismuth, selenium, tellurium, and tin, replacing traditional zinc metal and sodium borohydride reducing agents. The accessory simplifies the overall procedure and improves analysis efficiency.


Agilent VGA-76 Vapor Generation Accessory Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
100 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
702EAA979B021731CD32300DD3D29026

About this manual

The document introduces the Agilent VGA-76 automated vapor generation accessory, which utilizes a continuous flow technique with sodium borohydride as a reducing agent for atomic absorption analysis. It is designed to be connected to Agilent programmable sample changers and compatible with various Agilent atomic absorption spectrophotometers.

The note details the determination of specific elements including arsenic, antimony, bismuth, selenium, tellurium, tin, and mercury. It provides specific operating conditions, such as wavelengths and lamp currents, along with preparation procedures for calibration standards and reagents required for accurate analysis.

What's inside

  • Introduction
  • Instrumentation
  • Reagents and Samples
  • Design Features
  • Mercury Determination
  • Arsenic Determination
  • Antimony Determination

Specifications

Compatible Sample ChangerAgilent PSC-55
Integration period3 s
Typical delay timeabout 40 s
Mercury Wavelength253.7 nm
Mercury SBW0.5 nm
Mercury Lamp current4 mA
Arsenic Wavelength193.7 nm
Arsenic SBW0.2 nm
Arsenic Lamp current10 mA
Antimony Wavelength217.6 nm
Antimony SBW0.2 nm
Antimony Lamp current10 mA

Frequently asked questions

What elements can be determined using the VGA-76?

Elements that can be determined include arsenic, antimony, bismuth, selenium, tellurium, tin, and mercury.

What reducing agent is used in the VGA-76 system?

The system uses sodium borohydride as the reducing agent for the production of hydride vapor.

Can the VGA-76 be connected to an automated sample changer?

Yes, the accessory can be readily connected to an Agilent programmable sample changer to provide automatic presentation of up to 67 samples plus up to 5 calibration standards.

What is the detection limit for mercury with the flow-through cell?

A detection limit of 0.05 µg/L has been found with the flow-through cell for mercury.

Download PDF (100 KB)