Agilent SpectrAA-40 Application Note for Determination of Cu, Zn, Fe, Ca, Mg, Na, K in Serum

This article introduces a simple method for the determination of copper, zinc, iron, calcium, magnesium, sodium, and potassium in serum using flame atomic absorption spectroscopy. One milliliter of serum is diluted to 10 milliliters with deionized water, and the determination of seven elements is automatically performed.


Agilent SpectrAA-40 Application Note for Determination of Cu, Zn, Fe, Ca, Mg, Na, K in Serum - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
36 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
7E35083252343C7B9B4A49756966085B

About this manual

This document presents a simplified procedure for the determination of Copper (Cu), Zinc (Zn), Iron (Fe), Calcium (Ca), Magnesium (Mg), Sodium (Na), and Potassium (K) in serum. The method utilizes a SpectrAA-40 atomic absorption spectrometer with a programmable sample changer to automate the process, reducing the need for complex dilutions and anti-interference reagents compared to traditional methods.

The note details the specific instrument parameters, such as wavelength, slit width, and burner height, for each element. It also outlines the preparation of standard solutions and the procedure for analyzing serum samples diluted with deionized water. Results regarding recovery rates and precision are provided, along with a discussion on the effects of different dilution ratios and the comparison of this method against others.

What's inside

  • Introduction
  • Instrumentation
  • Parameters
  • Reagents
  • Standard Solutions
  • Procedure
  • Results
  • Discussion
  • References

Specifications

Instrument ModelSpectrAA-40
Sample ChangerPSC-56
PrinterEpson LX-80
Cu Wavelength324.8 nm
Zn Wavelength213.9 nm
Fe Wavelength248.3 nm
Ca Wavelength422.7 nm
Mg Wavelength202.5 nm
Na Wavelength330.3 nm
K Wavelength404.4 nm
Acetylene Flow13.5 L/min
Air Flow13.5 L/min
Measurement Time150 measurements in one hour

Frequently asked questions

What elements can be determined in serum using this method?

The method determines Copper (Cu), Zinc (Zn), Iron (Fe), Calcium (Ca), Magnesium (Mg), Sodium (Na), and Potassium (K).

What is the recommended dilution ratio for serum analysis?

One milliliter of serum is diluted to 10 milliliters with deionized water.

How long does the automatic measurement process take?

The procedure can be completed automatically with 150 measurements in one hour.

What instrument is used for this analysis?

The analysis uses a SpectrAA-40 atomic absorption spectrometer with a PSC-56 programmable sample changer.