Agilent Liberty 200 ICP-OES Application Note for Determination of Metals in Oils

This application note describes the technique for determining metals in oil samples using ICP-OES.


Agilent Liberty 200 ICP-OES Application Note for Determination of Metals in Oils - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
176 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
2FEFC1EE0540F1D435D15B45DA6899FC

About this manual

The document details an analytical procedure for determining twelve specific metals (Nickel, Barium, Silicon, Iron, Chromium, Magnesium, Vanadium, Sodium, Aluminium, Calcium, Zinc, and Lead) in various petroleum products. The study utilized the Agilent Liberty 200 ICP-OES instrument with a Compaq 386/20e computer for control.

Samples were prepared using a simple weight-volume dilution method with deodorized kerosene as the solvent, resulting in a 1:10 dilution ratio. The analysis employed manual sample introduction, dynamic background correction, and optimized viewing heights to ensure accurate results despite potential spectral interferences from hydrocarbon matrices.

What's inside

  • Introduction
  • Instrumentation
  • Analytical Procedures
  • Results
  • Measurement Parameters
  • Recovery Data
  • Conclusion

Frequently asked questions

What is the maximum number of replicates used for each sample analysis?

The study analyzed each sample in triplicate.

How long does it take to analyze 12 metals with three replicates?

The Liberty could analyze the 12 metals with 3 replicates in approximately 6 minutes 30 seconds.

What rinse time is required between fuel oil samples to minimize carry over?

A 3 minute rinse was used for the fuel oil samples.

Download PDF (176 KB)