Agilent 500 Ion Trap LC/MS Application Note for EPA 535

The performance of the Agilent 500 Ion Trap LC/MS in EPA Method 535 is evaluated in terms of calibration and quantitative accuracy in a variety of matrices. The data reveals this instrument provides reliable quantitative data, even with the presence of highly concentrated surface water extracts.


Agilent 500 Ion Trap LC/MS Application Note for EPA 535 - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
80 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
C81FF470DA5E457C24A1B344C5C57636

About this manual

The document evaluates the performance of the Agilent 500 Ion Trap LC/MS in EPA Method 535, specifically focusing on the determination of herbicide degradates (ethanesulfonic acids and oxanilic acid) in drinking water. The study assesses calibration and quantitative accuracy using surrogate standards in various matrices, including reagent water, finished tap water, and surface water extracts.

The analysis utilizes LC/MS/MS with electrospray ionization (ESI) negative mode. The note details the specific instrumentation setup, including the Agilent ProStar 210 Binary Solvent Delivery Modules and ProStar 420 AutoSampler. It provides specific HPLC gradient conditions and MS parameters, demonstrating that the instrument provides reliable quantitative data even with highly concentrated surface water extracts containing dissolved organic carbon.

What's inside

  • Abstract
  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • HPLC Conditions
  • MS Parameters
  • Results and Discussion
  • Conclusion
  • References

Specifications

Transition (Surrogate)300 > 121 m/z
API drying gas temp400 °C
API drying gas pressure35 psi
Scan range119-123 m/z

Frequently asked questions

What instrument is used for EPA Method 535 in this study?

The study evaluates the Agilent 500 Ion Trap LC/MS equipped with an electrospray ionization source (ESI) and built-in syringe pump.

What surrogate standard was used for matrix recovery and calibration?

The surrogate standard used was Dimethachlor ESA.

What is the correlation factor for the calibration curve?

The standard calibration curve for SS-Dimethachlor-ESA showed an excellent correlation factor (r²) value of 0.9999.

Did the surrogate recoveries meet EPA 535 method specifications?

Yes, the recovery criteria were between 70-130%, which is in agreement with the EPA 535 method.