This note describes an MSn data acquisition method using the 500 Ion Trap LC/MS TurboDDS software for the analysis of 7-ethoxycoumarin and two of its metabolites. TurboDDS saves time and streamlines data reporting for multiple stages of MS/MS and MSn analys.
This document details a data acquisition and analysis method using the Agilent 500 Ion Trap LC/MS equipped with TurboDDS software. The method focuses on the analysis of 7-ethoxycoumarin and two of its metabolites: 7-hydroxycoumarin and 7-hydroxycoumarin B-D-glucuronide. The study utilizes a specific HPLC gradient and mass spectrometry parameters to demonstrate the software's efficiency.
| Injection solvent | 50:50 methanol:water |
|---|---|
| API drying gas | 25 psi at 400 °C |
| API nebulizing gas | 25 psi |
| Needle voltage | 5000 V |
| Capillary voltage | 74AV |
| Shield voltage | 500 V |
| Enhanced scan mode | 5000 Da/sec |
| RF loading | 72% |
The TurboDDS software is used for the analysis of 7-ethoxycoumarin and its metabolites.
TurboDDS saves time and streamlines data reporting for multiple stages of MS/MS and MSn analysis.