Agilent Gradient Purification of Synthetic Acyl Carrier Protein Fragment 65-74 Application Note

This application note describes the use of Agilent PLRP-S 100Å column for small scale high efficiency separation of synthetic protein fragment.


Agilent Gradient Purification of Synthetic Acyl Carrier Protein Fragment 65-74 Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
572 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
ED271F06EE0C57B183449724D17A150D

About this manual

The document describes standard methods for small-scale peptide purification, specifically focusing on the separation of Acyl Carrier Protein (ACP) fragments. It highlights the use of PLRP-S 100A, a rigid polystyrene reversed phase media, to achieve high-efficiency separations for quantities in the micro-gram range.

The note details specific chromatographic conditions used for the separation, including column dimensions, eluent composition, gradient profile, flow rate, and detection wavelength. Results show the successful identification of ACP components and truncated sequences.

What's inside

  • Introduction
  • Conditions
  • Results and Discussion
  • Peak Identification

Specifications

Column Part NumberPL1512-3800
Media TypePLRP-S 100A
Particle Size8 um
Column Dimensions150 x 4.6 mm
Eluent A Composition0.1% TFA in 1 % ACN/99% water
Eluent B Composition0.1% TFA in 99 % ACN/1% water
Gradient Profile10-60% B in 20 min
Flow Rate1.0 mL/min
Detection WavelengthUV, 220 nm

Frequently asked questions

What is the purpose of this document?

To demonstrate the gradient purification of a synthetic acyl carrier protein fragment.

Which column media is used for this separation?

PLRP-S 100A, a rigid polystyrene (divinylbenzene) reversed phase media.

What are the eluent compositions?

Eluent A is 0.1% TFA in 1 % ACN/99% water and Eluent B is 0.1% TFA in 99 % ACN/1% water.