Agilent High Resolution Analysis of Triglycerides in Cooking Oil using HPLC with ELSD Application Note

This document describes the application of HPLC with ELSD for high-resolution analysis of triglycerides in cooking oil. ELSD detection offers higher sensitivity and adaptability compared to RI and UV, enabling the detection of low levels of contaminants. The author used a PLRP-S 100Å column and Agilent's ELSD detector for the analysis, and presented a chromatogram of unadulterated cooking oil.


Agilent High Resolution Analysis of Triglycerides in Cooking Oil using HPLC with ELSD Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
598 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
6B79AF498D52DD8A7E63AD8F58E6BB11

About this manual

The document details a method for analyzing the complex composition of triglycerides in cooking oil. It explains that gradient elution is required for high resolution separation and that Evaporative Light Scattering Detection (ELSD) is preferred over RI or UV due to solvent interference issues. The note highlights the suitability of PLRP-S 100A columns for this analysis due to their small pore sizes and high surface areas.

Specific instrumentation and conditions are provided, including the use of an Agilent ELS detector with specific settings (neb=40 °C, evap=40 °C, gas=1.6 SLM) and a PLRP-S 100A 5 um column. The method utilizes a gradient of Acetonitrile and Dichloromethane with a flow rate of 0.5 mL/min and an injection volume of 10 pL.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Sample Preparation
  • Conditions
  • Results and Discussion
  • Conclusion

Specifications

ColumnPLRP-S 100A 5 um, 150 x 4.6 mm
DetectorAgilent ELSD
Nebulizer Temperature40 °C
Evaporator Temperature40 °C
Gas Flow1.6 SLM
Flow Rate0.5 mL/min
Injection Volume10 pL

Frequently asked questions

Why is ELSD preferred over RI or UV for this analysis?

ELSD is unaffected by the optical properties of the solvents used in the gradient, whereas RI is not viable and UV can suffer from baseline disturbance.

What are the eluents used in the sample preparation?

Eluent A is Acetonitrile and Eluent B is Dichloromethane.

What is the sample concentration for the analysis?

The sample concentration is 1 mg oil/mL.