Polystyrene Oligomer Analysis on Agilent PLgel with Gel Permeation Chromatography Application Note

This application note describes the use of Agilent PLgel columns for the separation of low molecular weight polystyrene standards by gel permeation chromatography.


Polystyrene Oligomer Analysis on Agilent PLgel with Gel Permeation Chromatography Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
57 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
4A16F5CE69A20FDD046D3A11E3F4BB0C

About this manual

High performance, low pore size Agilent PLgel columns are ideal for resolving low molecular weight polystyrene standards into their oligomeric components. The document illustrates how successive peaks increase in molecular weight by the styrene repeat unit of 104.

The analysis utilizes a system consisting of two Agilent PLgel 5 µm 100A columns (600 x 7.6 mm) with THF as the eluent, a flow rate of 1.0 mL/min, and an RI detector.

What's inside

  • Introduction
  • Structures of low molecular weight polystyrene oligomers
  • Analysis Conditions
  • Conclusion

Frequently asked questions

What is the molecular weight of the lowest peak assigned using phenyl hexane as a marker?

The lowest MW peak can be assigned using phenyl hexane as a marker, with a MW of 162.

What is the molecular weight increase for successive peaks in polystyrene oligomers?

Successive peaks increase in molecular weight by the MW of the styrene repeat unit, 104.

What is the flow rate used for the analysis on an Agilent PLgel column set?

The flow rate used for the analysis is 1.0 mL/min.