Agilent PLgel 3 µm 100A Dialkyl Phthalate Analysis Application Note

This article describes the analysis of dialkyl phthalates using gel permeation chromatography on Agilent PLgel 3 µm columns. The method is suitable for analyzing low molecular weight polymer additives such as dialkyl phthalates.


Agilent PLgel 3 µm 100A Dialkyl Phthalate Analysis Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
43 KB
Published
11 June, 2012
Updated
29 September, 2023
MD5 checksum
6ED9FE61C03DEF191992FA954CFDA3E1

About this manual

This document details the analysis and quantitation of dialkyl phthalates, which act as plasticizers in commercial polymers. Agilent PLgel 3 µm 100A columns are specifically designed to exclude polymers with molecular weights over 4,000 g/mol, allowing for the resolution of low molecular weight additives.

What's inside

  • Introduction
  • Analysis of Dialkyl Phthalates
  • Conditions
  • Conclusions

Specifications

Column dimensions300 x 7.6 mm
Column part numberPL1110-6320
EluentTHF
Flow rate1.0 mL/min
Injection volume20 pL
Detector typeRI
System modelAgilent PL-GPC 50

Frequently asked questions

What concentration of dialkyl phthalates is used in the sample preparation?

The analyte samples were prepared at a concentration of 0.2% (w/v) in tetrahydrofuran.

What is the flow rate used for the analysis?

The analysis is performed with a flow rate of 1.0 mL/min.

What injection volume is required for the procedure?

The samples are injected with a volume of 20 pL.