Quantitative Analysis of Vitamin C by Ion Suppression Chromatography Application Note

This application note describes the optimization of an ion suppression reversed phase HPLC method to detect L-ascorbic acid, and use of the method in the quantitative determination of the acid in fruit juice. PLRP-S columns were used in this investigation. PLRP-S is a rigid macroporous styrene/divinylbenzene HPLC phase with outstanding chemical and physical stability, and high surface area. This chemical stability permits the use of low pH eluents and ion suppression reversed phase HPLC.


Quantitative Analysis of Vitamin C by Ion Suppression Chromatography Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
756 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
D84F68906D8EDEE48F333DE90E187CB3

About this manual

This document details the optimization of an ion suppression reversed phase High Performance Liquid Chromatography (HPLC) method to detect and quantify L-ascorbic acid. The study utilizes PLRP-S columns, which are described as rigid macroporous styrene/divinylbenzene phases with high chemical stability allowing for low pH eluents.

The optimized conditions involve a mobile phase of 0.2 M sodium dihydrogen phosphate adjusted to pH 2.14 with HCl. The method successfully resolves L-ascorbic acid from D-erythorbic acid and the solvent front within nine minutes, enabling accurate quantification in fruit juices without derivatization.

What's inside

  • Introduction
  • Materials and Reagents
  • Conditions
  • Results and Discussion
  • Conclusion

Frequently asked questions

What is the purpose of this application note?

This document describes how to optimize an ion suppression reversed phase HPLC method to detect L-ascorbic acid and use it for quantitative determination in fruit juice.

What are the dimensions of the column used?

The investigation uses a PLRP-S column with dimensions of 250 x 4.6 mm.

What is the flow rate for the analysis?

The specified flow rate for the method is 0.5 mL/min.