Agilent PLgel 10 MIXED-B GPC/SEC Application Note for PVDF Analysis

This document discusses the analysis of polyvinylidene fluoride (PVDF) on Agilent PLgel 10 µm MIXED-B and GPC/SEC. PVDF is an important material with a wide range of uses from non-stick coatings to inert packaging materials. However, PVDF can be difficult to analyze by gel permeation chromatography due to its low solubility in common solvents. This challenge can be overcome by using dimethyl sulfoxide as a solvent, Agilent PLgel MIXED-B columns, and operating the GPC system at elevated temperature with appropriate high temperature sample preparation. In this study, three samples of PVDF were analyzed in dimethyl sulfoxide at 95 °C using the Agilent PL-GPC 220 instrument with two Agilent PLgel 10 µm MIXED-B, 300 × 7.5 mm columns. By operating the GPC system at a flow rate of 1.0 mL/min and heating the samples for 4 hours to dissolve them and reduce the solvent viscosity, satisfactory chromatographic separation was achieved.


Agilent PLgel 10 MIXED-B GPC/SEC Application Note for PVDF Analysis - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
53 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
2E787C1BE6FC94C0D8AB3D22E4EFCD1E

About this manual

This document details a method for analyzing polyvinylidene fluoride (PVDF) using gel permeation chromatography (GPC). It addresses the difficulty of analyzing fluoropolymers due to their low solubility in common solvents. The recommended approach involves using dimethyl sulfoxide as a solvent, Agilent PLgel MIXED-B columns, and operating the system at 95 °C.

What's inside

  • Introduction
  • Analysis of Polyvinylidene Fluoride
  • Conditions
  • Conclusions

Specifications

Columns2 x Agilent PLgel 10 MIXED-B, 300 x 7.5mm
EluentDimethyl sulfoxide
Flow rate1.0 mL/min
Injection volume100 pL
Temperature95 °C
SystemPL-GPC 220

Frequently asked questions

What solvent is recommended for analyzing PVDF?

Dimethyl sulfoxide is recommended to overcome the low solubility of fluoropolymers in common solvents.

What temperature is required for the analysis?

The system must be operated at 95 °C to dissolve samples and reduce solvent viscosity.