Agilent HPLC Analysis of Epikote 828 & 1001 Application Note

This application note describes the conditions and results for the separation of Epikote 828 and 1001 using an Agilent HPLC system.


Agilent HPLC Analysis of Epikote 828 & 1001 Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
668 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
BDC4F7E92B4B4C8983701C301187A388

About this manual

This document details the High-Performance Liquid Chromatography (HPLC) analysis of Epikote 828 and Epikote 1001 epoxy resins. It explains that PLRP-S, a rigid macroporous styrene/divinylbenzene phase, is ideal for assessing the oligomeric composition of these resins due to its chemical stability and robustness across various eluents.

The note provides specific analytical conditions, including a sample concentration of 1.3 mg/mL, a PLRP-S column (p/n PL1512-5801), and a specific gradient elution profile using water, acetonitrile, and tetrahydrofuran. Results indicate that Epikote 828 contains greater quantities of smaller oligomers compared to Epikote 1001, which contains larger quantities of higher oligomers.

What's inside

  • Introduction
  • Conditions
  • Results and Discussion
  • Conclusion

Specifications

Sample Concentration1.3 mg/mL
ColumnPLRP-S 300A 8 ym, 250 x 4.6 mm
Column Part NumberPL1512-5801
Eluent A100% water
Eluent B90% ACN, 10% THF
Detection Wavelength254 nm

Frequently asked questions

Which column is recommended for analyzing Epikote resins?

The PLRP-S column is recommended for the assessment of the oligomeric composition of epoxy resins.

What is the difference between Epikote 828 and 1001 in terms of oligomers?

Epikote 828 contains greater quantities of smaller oligomers, while Epikote 1001 contains larger quantities of higher oligomers.

Can the PLRP-S column switch between different organic modifiers?

Yes, it is possible to switch between organic modifiers such as methanol and tetrahydrofuran.

Download PDF (668 KB)