Agilent 1290 Infinity LC Application Note: UV MS Detection Sensitivity

This document introduces the characteristics of Agilent 1290 Infinity LC and its complementary sensitivity with MS detection.


Agilent 1290 Infinity LC Application Note: UV MS Detection Sensitivity - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
470 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
1558C61481C3B5701B2AC7B6B782673B

About this manual

This document evaluates the complementary sensitivity of UV and MS detection in fast LC/MS analysis. It compares the limits of detection (LOD) and quantification (LOQ) achieved by an Agilent 6140 Single Quadrupole Mass Spectrometer against different Diode Array Detector (DAD) configurations on the Agilent 1290 Infinity LC system.

The study specifically analyzes the performance of the Agilent 1290 Infinity DAD equipped with 10-mm and 60-mm Max-Light cartridge cells, as well as a comparison to the Agilent 1200 Series DAD SL. Results indicate that the 60-mm Max-Light cell configuration offers the highest sensitivity among the DAD setups, narrowing the gap between UV and MS detection capabilities.

What's inside

  • Introduction
  • Experimental
  • Results and discussion
  • Calibration curves
  • Limits of detection and quantification
  • Precision of peak areas
  • Conclusion

Specifications

System ComponentsAgilent 1290 Infinity Binary Pump, Autosampler, Thermostatted Column Compartment, DAD, Agilent 6140 Single Quadrupole Mass Spectrometer
Flow Rate1.0 mL/min
Column Temperature40°C
MS Source Temperature350 °C
MS Nebulizer Pressure60 psi
MS Capillary Voltage4000 V
MS PolarityPositive
MS SIM Peak Width0.04 min
MS SIM Cycle Time0.24 s
MS SIM Gain1.0
MS SIM m/z 279.10120 V, 114 msec dwell
MS SIM m/z 311.10120 V, 114 msec dwell
DAD 1200 Series SL LOQ200 pg/pL

Frequently asked questions

Which DAD configuration provides the best sensitivity on the Agilent 1290 Infinity?

The Agilent 1290 Infinity DAD equipped with the 60-mm Max-Light cartridge cell provides the best overlap with mass spectrometric detection.

How does the sensitivity of the 60-mm Max-Light cell compare to the 10-mm cell?

The 60-mm cell provides a 5 times higher sensitivity compared to the 10-mm Max-Light cartridge cell.

What is the limit of quantification for the Agilent 1200 Series DAD SL?

The Agilent 1200 Series DAD SL with a 10-mm cell has a limit of quantification of 200 pg/pL.

Download PDF (470 KB)