Agilent 490 Micro GC Application Note for Tetrahydrothiophene Analysis

The application note shows the analysis of Tetrahydrothiophene (THT) in a Natural Gas matrix using the Agilent 490 Micro GC.


Agilent 490 Micro GC Application Note for Tetrahydrothiophene Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
537 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
4D74FFEBCB7035DA26B39F87790031C2

About this manual

This document details an application for analyzing Tetrahydrothiophene (THT), used as an odorant in natural gas, utilizing the Agilent 490 Micro GC. The method employs a specific CP-Sil 19 CB column channel optimized to separate THT from other compounds like Nonane.

What's inside

  • Introduction
  • Instrumentation

Specifications

Column temperature90 °C
Carrier gasHelium, 200 kPa
Injector temperature110 °C
Injection time255 msec
Sampling time30 sec

Frequently asked questions

What is the total analysis time for THT?

The total analysis time is less than 90 seconds.

When does Tetrahydrothiophene elute?

Tetrahydrothiophene elutes around 40 seconds.