Agilent Mine Gas Analysis Using the 490-GC Micro-GC Manual

This application note describes the use of the 490 Micro GC for the analyses of mine gas, especially carbon monoxide at low level, detection of which is essential for the safety of mine workers. Three independent GC channels analyze the sample in less than two minutes, including permanent gases and C1-C2 hydrocarbons.


Agilent Mine Gas Analysis Using the 490-GC Micro-GC Manual - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
640 KB
Published
11 June, 2012
Updated
30 September, 2023
Model
490-GC
MD5 checksum
4DA32683F8A44B2009062AA5ED8B4CDE

About this manual

This document details a method for analyzing mine gases using the Agilent 490 Micro Gas Chromatograph (Micro GC). The primary focus is on the detection of trace carbon monoxide (CO) at low concentrations within bulk methane, which is critical for mine worker safety. The analysis utilizes three independent channels to separate and detect permanent gases, C1-C2 hydrocarbons, and other components in less than two minutes.

The procedure specifies distinct chromatographic conditions for each channel. Channel 1 uses argon as a carrier gas on a Molsieve column to measure hydrogen, helium, and other permanent gases. Channel 2 employs helium with an extra-high sensitivity mode to detect very low levels of carbon monoxide without interference from methane. Channel 3 analyzes carbon dioxide, ethane, and ethylene using a CP-PoraPLOT Q column.

What's inside

  • Instrumentation
  • Materials and Methods
  • Chromatographic conditions
  • Results and Discussion
  • Peak identification

Frequently asked questions

What is the primary safety application of this instrument?

The instrument is used to detect carbon monoxide at low levels, which is essential for mine worker safety.

How long does it take to analyze a sample?

Three independent channels analyze the sample in less than two minutes.

What is the Limit of Quantification for oxygen, nitrogen, and carbon monoxide?

The limit is approximately 50 to 100 ppm for these gases.