Quantitative end-point RT-PCR gene expression measurement using the Agilent 2100 Bioanalyzer and standardized RT-PCR

This application note describes how the Agilent 2100 bioanalyzer can be used in combination with Standardized RT (StaRT)-PCR reagents from Gene Express National Enterprises (G.E.N.E.), Inc. (Huntsville, AL, USA) to obtain highly reproducible, standardized gene expression measurement with RT-PCR. The internal standards used in StaRT-PCR control for variable loading of PCR products into electrophoresis chan- nel, and allow quantitative gene expression measurement at the end- point of PCR, without real-time analysis. Gene expression measure- ment of 15 genes was more reproducible with the Agilent 2100 bioana- lyzer than with the ABI Prism 310 Genetic Analyzer or agarose gel elec- trophoresis methods. The mean coefficient of variance (CV) for these three methods was 0.26, 0.39 and 0.50, respectively.


Quantitative end-point RT-PCR gene expression measurement using the Agilent 2100 Bioanalyzer and standardized RT-PCR - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
716 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
348D0C79A03096900454A32E80785992

About this manual

The document details how the Agilent 2100 Bioanalyzer can be used in combination with Standardized RT (StaRT)-PCR reagents from Gene Express National Enterprises to obtain highly reproducible, standardized gene expression measurements. The internal standards used in StaRT-PCR control for variable loading of PCR products into electrophoresis channels.

The study compares the reproducibility of this method against agarose gel electrophoresis and the ABI Prism 310 Genetic Analyzer. Results indicate that gene expression measurement of 15 genes was more reproducible with the Agilent 2100 bioanalyzer, showing a mean coefficient of variance (CV) of 0.26 compared to 0.39 for the ABI Prism 310 and 0.50 for agarose gel methods.

What's inside

  • Abstract
  • Introduction
  • StaRT-PCR Protocols
  • Electrophoresis and Quantitation
  • Agilent 2100 Bioanalyzer Analysis
  • Agarose Gel Electrophoresis
  • Capillary Electrophoresis
  • Statistical Analysis
  • Results Comparison

Frequently asked questions

What is the mean coefficient of variance (CV) for gene expression measurement using the Agilent 2100 bioanalyzer?

The mean CV was 0.26, which was lower than the ABI Prism 310 Genetic Analyzer (0.39) and agarose gel electrophoresis methods (0.50).

Which reagents are used with the Agilent 2100 bioanalyzer in this method?

The method uses Standardized RT (StaRT)-PCR reagents from Gene Express National Enterprises.