Agilent Bond Elut DMS Card Application Note: Homogeneity of Dried Matrix Spots

This Application Note discusses the concept of dried matrix spot homogeneity and demonstrates the homogeneity of blood spots on an Agilent Bond Elut DMS card. Results show that the size of the blood spots and the analyte dispersion is homogeneous, and the recoveries are reproducible regardless of whether the sample is taken from the center or the edge of the blood spot.


Agilent Bond Elut DMS Card Application Note: Homogeneity of Dried Matrix Spots - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
62 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
19C1F96BBB8CE1E95172CF3B78D68215

About this manual

Dried blood spotting is an alternative to traditional plasma extraction for toxicokinetic and pharmacokinetic data. This document investigates whether the size of the blood spot and the dispersion of analytes are consistent across different locations within the spot.

Four pharmaceutical analytes (clozapine, zolpidem, paroxetine, and nortriptyline) were tested using 20 µL drops of spiked human blood on Agilent Bond Elut DMS cards. The results demonstrated that spot diameters had an RSD of 3% or lower across three different lots, and analyte recoveries were consistent regardless of whether the sample was punched from the center or the edge.

What's inside

  • Introduction
  • Experimental
  • LC/MS conditions
  • Results and Discussion
  • Conclusions

Frequently asked questions

What is the maximum RSD value for spot diameter across different lots?

The RSD values for the diameter of the spots were all 3% or lower.

Are Agilent Bond Elut DMS cards suitable for diagnostic procedures?

No, they are intended only for DMPK/ADME research applications.