Agilent Application Note: Determination of Calcium and Magnesium in Blood Serum and Urine

This application note describes a method for the determination of calcium and magnesium in blood serum and urine using atomic absorption spectrophotometry.


Agilent Application Note: Determination of Calcium and Magnesium in Blood Serum and Urine - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
138 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
509EDFCD7F2973652000E298D86332C4

About this manual

This document outlines the application of atomic absorption spectrophotometry for analyzing calcium and magnesium in clinical materials, specifically blood serum and urine. It details sample preparation methods, including dilution factors and the use of reagents like lanthanum chloride or strontium chloride to suppress interferences from proteins and phosphates.

The note provides specific instrumental settings for the Techtron Model 1000 atomic absorption spectrophotometer, including resonance lines, lamp currents, and burner types. It also discusses the effects of flame type (air-acetylene vs. nitrous oxide-acetylene) on sensitivity and includes calibration data.

What's inside

  • Introduction
  • Determination of Calcium
  • Determination of Magnesium
  • Instrumentation
  • Sample Solutions
  • Standard Solutions
  • Results and Discussion
  • Conclusion

Specifications

Resonance Line (Calcium)4226.7 A
Resonance Line (Magnesium)2852.1 A
Photomultiplier (Calcium)R-213
Lamp Current (Calcium)4mA
Lamp Current (Magnesium)3mA
Spectral Band Pass2 A
Flame Type (Calcium)N2O-C,H Air-C,H
Burner (Calcium)AB-50, 5 cm
Burner (Magnesium)AB-51, 10 cm
Sample Volume (Serum)0.5 mL
Sample Volume (Urine)0.5 mL

Frequently asked questions

What flame is recommended for determining calcium?

The use of the nitrous oxide-acetylene flame is recommended for the determination of calcium at the 4226.7 A resonance line.

What flame is used for determining magnesium?

Magnesium is determined in the air-acetylene flame at the 2852.1 A resonance line.

What role does lanthanum play in the solutions?

The lanthanum present in the solutions will control interference effects and act as an ionization suppressant.

How much does sensitivity decrease if an air-acetylene flame is used for calcium instead of nitrous oxide-acetylene?

If nitrous oxide is not available, an air-acetylene flame may be employed, but the sensitivity is approximately 3X less.

Download PDF (138 KB)