Agilent 6550 Q-TOF High Sensitivity Peptide Analysis Application Note

This document describes the application of the Agilent 6550 Quadrupole Time-of-Flight (Q-TOF) Mass Spectrometer for high sensitivity peptide analysis. The Agilent iFunnel technology combines highly efficient electrospray ion generation and focusing with a hexabore capillary sampling array and dual-stage ion funnel to increase ion sampling and transmission. This innovative design allows for a 10-fold increase in sensitivity for selected molecules compared to an earlier generation Q-TOF instrument without iFunnel technology.


Agilent 6550 Q-TOF High Sensitivity Peptide Analysis Application Note - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
664 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
0CA56C7AF9BC2B64F368DD15DB0BBCFC

About this manual

This document describes the high sensitivity capabilities of the Agilent 6550 Quadrupole Time-of-Flight (Q-TOF) Mass Spectrometer equipped with iFunnel technology. The technology combines efficient electrospray ion generation with a hexabore capillary sampling array and dual-stage ion funnel to increase ion transmission. The note highlights that this design allows for a 10-fold increase in sensitivity for selected molecules compared to earlier generation Q-TOF instruments lacking iFunnel technology.

The document presents experimental data comparing the Agilent 6550 iFunnel O-TOF system against an Agilent 6540 O-TOF system. Results show significant increases in signal height for a synthetic peptide in both full scan MS and targeted MS/MS modes. Additionally, the note details the system's performance in protein identification workflows and peptide quantitation, demonstrating the ability to identify proteins at the 10 amol on-column level and achieve linearity over four orders of dynamic range.

What's inside

  • Introduction
  • Improved Sensitivity with iFunnel Technology
  • Improved Sensitivity for Protein Identification
  • Improved Sensitivity for Peptide Quantitation
  • Conclusions

Specifications

Sensitivity increase over earlier generation O-TOF10-fold increase
Sensitivity increase in targeted MS/MS mode7x increase
Protein identification limit10 amol on-column
Peptide quantitation limit of detection10 amol on-column
Dynamic range for quantitation4 orders

Frequently asked questions

How much does iFunnel technology improve sensitivity compared to an earlier generation O-TOF?

It enables a 10-fold increase in sensitivity for selected molecules.

What is the limit of detection for peptide quantitation using the 6550 iFunnel O-TOF?

The limit of detection is 10 amol on-column for the target peptide in the presence of 150 ng of digested yeast lysate.

Can the 6550 iFunnel O-TOF perform high sensitivity qualitative and quantitative analysis on a single system?

Yes, the system permits high sensitivity qualitative and quantitative analysis to be performed on a single system.