Agilent 4100 MP-AES Application Note for Diesel Analysis

This document describes the analysis of trace elements in petro-diesel and biodiesel fuels using the Agilent 4100 MP-AES. The instrument is capable of stable analysis of organic matrices and eliminates the need for expensive and dangerous gases, resulting in lower running costs and improved productivity.


Agilent 4100 MP-AES Application Note for Diesel Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
184 KB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
0BF6608B420CC4A95B6956086199E710

About this manual

The document details a method for analyzing trace elements (Mg, Ca, Na, K) in petro-diesel and biodiesel to prevent engine corrosion. It highlights how the Agilent 4100 MP-AES uses magnetically-coupled microwave energy with nitrogen gas to create a stable plasma capable of handling challenging organic matrices without expensive or dangerous gases like acetylene.

The procedure utilizes an External Gas Control Module (EGCM) and the OneNeb nebulizer to prevent carbon deposition and reduce background emissions. The note provides specific instrument operating conditions, analyte settings for wavelengths and pressures, method detection limits in parts per billion, and results from spike recovery tests demonstrating accuracy within +10% of target values.

What's inside

  • Introduction
  • Instrumentation
  • Instrument Operating Conditions
  • Analyte Settings
  • Samples and Sample Preparation
  • Detection Limits
  • Spike Recoveries
  • Conclusions

Specifications

Nebulizer TypeInert OneNeb
Spray Chamber TypeDouble-pass glass cyclonic
Operating SystemMicrosoft Windows 7
Mg Detection Limit (MDL)27 ppb
Ca Detection Limit (MDL)8.2 ppb
Na Detection Limit (MDL)18.7 ppb
K Detection Limit (MDL)27 ppb

Frequently asked questions

What is the method detection limit for Magnesium (Mg) in this analysis?

The method detection limit for Mg is 27 ppb.

How does the External Gas Control Module (EGCM) improve analysis?

It allows air injection into the plasma to prevent carbon deposition and reduce background emissions.

Download PDF (184 KB)