Agilent Bond Elut Plexa SPE HPLC Application Note for Phenols Determination

This application note describes the development of a sample extraction, cleanup, and enrichment method with an Agilent Bond Elut Plexa cartridge and a separation method using an Agilent Poroshell 120 column, with quantification by HPLC. The method is simple and effective for the extraction, enrichment, and analysis of multiple phenol compounds in drinking water.


Agilent Bond Elut Plexa SPE HPLC Application Note for Phenols Determination - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
58 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
1F7692E2D67643DD3AB381A04B499F6A

About this manual

This document details a method for the simultaneous determination of 11 phenols in drinking water. The procedure utilizes solid phase extraction (SPE) with an Agilent Bond Elut Plexa cartridge for sample extraction, enrichment, and cleanup. Separation is achieved using an Agilent Poroshell 120 column with HPLC.

What's inside

  • Abstract
  • Introduction
  • Reagents and chemicals
  • Standard solutions
  • Equipment and materials
  • HPLC conditions
  • Sample preparation
  • Results and Discussion
  • Conclusion

Specifications

Solid phase extraction cartridgeAgilent Bond Elut Plexa cartridge, 80 mg, 3 mL
Vacuum manifoldAgilent Vac Elut 20 Manifold Tall Glass Basin
Mobile phase A0.1% Formic acid in water
Mobile phase B0.1% Formic acid in acetonitrile
Flow rate0.4 mL/min
DetectionUV, 280 nm
Sample volume250 mL

Frequently asked questions

Which phenols are analyzed in this method?

The method analyzes 11 phenol compounds including Phenol, 4-nitrophenol, 2-chlorophenol, and Pentachlorophenol.

What is the recovery range for the phenols?

Overall recoveries ranged from 87% to 108%.

What is the RSD range for the analysis?

The RSD values were between 1.4% and 6.7%.

Download PDF (58 KB)

More Agilent Manuals


View all Agilent manuals →