Agilent Cary 630 FTIR Application Note for PE/PP Blend Analysis

This application note describes a method for rapidly determining the PE:PP ratio in blends using the Agilent Cary 630 FTIR spectrometer.


Agilent Cary 630 FTIR Application Note for PE/PP Blend Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
435 KB
Published
11 June, 2012
Updated
30 September, 2023
MD5 checksum
D5376C7FF089A2BD5B3A53261A876FDA

About this manual

The document details a method for rapidly determining the ratio of polyethylene (PE) to polypropylene (PP) in polymer blends using the Agilent Cary 630 FTIR spectrometer. It addresses the importance of knowing blend composition for performance and recycling applications.

Two procedures are described: a traditional salt plate cast film method consistent with ASTM D3900-05a, and a novel PTFE cast film method utilizing the DialPath or TumbllR accessory. Both methods utilize specific infrared absorbance ratios to calculate weight percent PE.

What's inside

  • Introduction
  • Experimental
  • Calibration standards
  • Salt plate cast film procedure
  • PTFE cast film procedure
  • DialPath and TumbllR accessories
  • Results and discussion
  • FTIR calibration plots
  • Conclusions

Specifications

Measurement time30 seconds
Resolution4 cm-1
Concentration range35-85% PE
Absorbance range0.3-1.0 AU

Frequently asked questions

Which Agilent instruments support the DialPath method?

The new DialPath method can be used on the Agilent 4500, 5500 FTIR spectrometers as well as the Agilent Cary 630 FTIR.

What is the calibration correlation coefficient for this method?

The new cast film method based on the DialPath accessory yields an R value of 1.000.