Agilent 1290 Infinity LC System Application Note for Primary Aromatic Amine Detection

This Application Note demonstrates that stringent sensitivity requirements for the detection of potentially harmful primary aromatic amines can be fulfilled when using an Agilent 1290 Infinity LC System equipped with the 1290 large volume injection kit. This enables the Agilent 1290 Infinity Autosampler for injection volumes up to 120 μL for sample enrichment on the column. A high sensitivity 60-mm Max-Light flow cell in the Agilent 1290 Infinity Diode Array Detector gives additional detection sensitivity.


Agilent 1290 Infinity LC System Application Note for Primary Aromatic Amine Detection - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
496 KB
Published
11 June, 2012
Updated
01 October, 2023
MD5 checksum
AC5D4621AA6CD25321249BE2FB726C19

About this manual

This document describes an analytical method developed on the Agilent 1290 Infinity LC System for the detection and quantification of primary aromatic amines (PAAs) in consumer products like printed paper and plastics. The method addresses legal requirements to detect these potentially harmful compounds at levels as low as 1 ppb.

What's inside

  • Abstract
  • Introduction
  • Equipment
  • Method
  • Results and discussion
  • Conclusion

Frequently asked questions

What is the maximum injection volume supported by the Agilent 1290 Infinity Autosampler when equipped with the large volume injection kit?

The system enables injection volumes up to 120 pL for sample enrichment on the column.

What is the limit of detection (LOD) found for primary aromatic amines using this method?

The limit of detection was found at 0.5 ppb.

What is the defined limit of quantification (LOQ) for individual primary aromatic amines in this method?

A concentration level of 2 ng/mL was defined as the limit of quantification.