Agilent Comprehensive Profiling of Free Conjugated Estrogens by CE-TOF-MS Application Note

This document describes a comprehensive method for profiling free and conjugated estrogens using capillary electrophoresis-time of flight-mass spectrometry. The method is sensitive and robust.


Agilent Comprehensive Profiling of Free Conjugated Estrogens by CE-TOF-MS Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
12
File format
PDF
File size
1.5 MB
Published
11 June, 2012
Updated
28 September, 2023
MD5 checksum
0C11C80ABCBF895C2FF1555BF21EE57C

About this manual

This document details a methodology for the comprehensive profiling of free and conjugated estrogens in complex biological fluids, such as urine. The method addresses the challenges of wide dynamic range and chemical diversity in endogenous and exogenous estrogen metabolites.

What's inside

  • Introduction
  • Chemicals and Reagents
  • Instrumentation
  • CE-ESI-TOF-MS Conditions
  • Experimental Design
  • Molecular Modeling
  • Urine Collection and Sample Pretreatment
  • Method Calibration and Assay Validation

Specifications

Background Electrolyte50 mM ammonium bicarbonate pH 9.5
Injection Pressure50 mbar
Injection Time5 seconds
Applied Voltage20 kV
Sheath Liquid Flow Rate14 pL/min
Mass Rangem/z 50-1100
Mass Accuracy<2 ppm

Frequently asked questions

What is the optimized capillary voltage for maximizing estrogen detection?

The optimum condition is a capillary voltage of -4.0 kV.

What is the sheath liquid composition for the optimized method?

The sheath liquid is 5 mM ammonium bicarbonate in 80:20 MeOH:H2O.