Agilent Bond Elut Polymer SCX SPE Application Note for Benzimidazole Fungicides

Solid-phase extraction (SPE) coupled with high-performance liquid chromatography (HPLC) was optimized for extraction and quantification of two fungicides (carbendazim and thiabendazole) in apple juice. Results indicate that SPE using Agilent Bond Elut SCX (60 mg, 3 mL) and HPLC using an Agilent ZORBAX Eclipse Plus C18 column (4.6 mm × 100 mm, 3.5 µm) is an excellent combination for extraction and analysis of these compounds. Recoveries ranged from 92.1 to 99.4 percent with RSDs below 5 percent and limits of detection of 4 µg/kg.


Agilent Bond Elut Polymer SCX SPE Application Note for Benzimidazole Fungicides - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
526 KB
Published
08 June, 2012
Updated
28 September, 2023
MD5 checksum
2A4379C7B4116EB43144F9E3E7065B9C

About this manual

This document details an optimized method for extracting and quantifying carbendazim and thiabendazole fungicides from apple juice. The procedure utilizes Agilent Bond Elut SCX cartridges coupled with High-Performance Liquid Chromatography (HPLC) using a ZORBAX Eclipse Plus C18 column.

The study validates the method based on SN/T 1753-2006, reporting recoveries between 92.1% and 99.4%, relative standard deviations below 5%, and limits of detection at 4 µg/kg.

What's inside

  • Introduction
  • Materials and Chemicals
  • HPLC Conditions
  • Sample Preparation
  • SPE Purification
  • Results and Discussion

Frequently asked questions

What is the limit of detection for carbendazim and thiabendazole in apple juice?

The limit of detection for both fungicides is 4 pg/kg.

What are the recovery percentages for spiked fungicide standards at 25, 50, and 100 pg/kg levels?

Recoveries for carbendazim ranged from 95.9% to 99.4%, while recoveries for thiabendazole ranged from 92.1% to 99.0%.

What is the required flow rate when loading the sample onto the SPE cartridge?

The sample should be loaded at a speed of about 1 mL/min.