Thermo Scientific KingFisher Flex Magnetic Particle Processor User Manual

The Thermo Scientific KingFisher Flex User Manual provides information on the product, including the installation and operating instructions.


Thermo Scientific KingFisher Flex Magnetic Particle Processor User Manual - cover page
Brand
Thermo Scientific
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
79
File format
PDF
File size
3.1 MB
Published
29 November, 2023
Updated
29 November, 2023
MD5 checksum
763B3F8FC88178D8E0869835F4E083FE

About this manual

This user manual describes the Thermo Scientific KingFisher Flex magnetic particle processor, an instrument intended for professional laboratory use by trained personnel for the automated transfer and processing of magnetic particles in microplate format. It is written for the end user, such as a research scientist or laboratory technician, and covers safety precautions, installation and operating instructions.

The document explains the principle of magnetic particle processing with magnetic rods and disposable tip combs, how to install and start up the instrument, how to run protocols with the keypad, display and internal software, and how to perform cleaning and preventive maintenance. It also contains a troubleshooting guide with error messages, a list of accessories, consumables and spare parts, and the general and performance specifications of the instrument.

What's inside

  • Safety symbols and markings
  • Instrument safety and guidelines for use
  • Introduction to the KingFisher Flex
  • Principle of operation
  • Installation
  • Routine operation
  • Maintenance
  • Technical specifications
  • Frequently asked questions
  • Troubleshooting guide

Specifications

Overall dimensions (instrument)ca. 680 mm (W) x 600 mm (D) x 380 mm (H) [26.8" (W) x 23.6" (D) x 15" (H)]
Weight (instrument)ca. 28 kg [62 lbs.]
Operating conditionsIndoor use, +5 °C to +40 °C; maximum relative humidity 80% for temperatures up to 31 °C, decreasing linearly to 50% relative humidity at 40 °C
Mains power supply100–240 Vac, 50/60 Hz, nominal; automatic voltage detection
Power consumption175 VA max.; 55 VA standby
Heat dissipation341 BTU max.
Computer interfaceUSB or serial RS-232C port; baud rate 9600; 1 start bit, 8 data bits, 1 stop bit, no parity; flow control XON/XOFF
Internal softwareSpace for max. 512 internal protocols
Transportation conditions-40 °C to +60 °C, packed in transport packaging

Frequently asked questions

What is the KingFisher Flex intended for?

It is intended for the automated transfer and processing of magnetic particles in microplate scale, for professional laboratory use by trained personnel.

What mains power supply does the instrument use?

The instrument uses 100–240 Vac, 50/60 Hz nominal, with automatic voltage detection.

Who should not use this product?

People wearing a pacemaker or metallic prostheses should not use the product, because it contains very strong permanent magnets.

How many protocols can be stored in the internal software?

There is space for a maximum of 512 internal protocols.