Baby Leo Hybrid Incuwarmer User Guide. This guide will introduce users how to effectively utilize the device and its functions, optimizing care for vulnerable populations.
This user guide explains how to power on, operate, and shut down the Baby Leo hybrid incuwarmer. It covers patient selection, thermoregulation modes, weighing, sound monitoring, observation lighting, height and tilt adjustment, and special procedures.
The guide also describes radiant warmer and incubator operation, including humidity, condensation, touch-time air curtain, heated mattress use, and the required shutdown sequence.
| Start process time | Approximately 7 seconds |
|---|---|
| Patient data retention after shutdown | Deleted after 2 hours |
| Background noise level in incubator mode | 40 dB |
| Noise alarm threshold | Greater than 55 dB |
| Heated mattress temperature range | 35-39 C |
| Heated mattress temperature increment | 0.5 C |
| Touch-time air curtain duration | 20 minutes |
| Radiant warmer prolonged high-output alarm | At 14 minutes; warmer off at 15 minutes |
The start process takes approximately 7 seconds.
Patient data is deleted after 2 hours.
The background noise level is 40 dB.
The heated mattress can be set from 35-39 C in increments of 0.5 C.