User guide for the Integra CUSA Excel and CUSA Excel+ ultrasonic surgical aspirator system, covering setup, operation, safety and maintenance.
The Integra CUSA Excel/CUSA Excel+ User’s Guide provides detailed instructions for qualified medical professionals on using the ultrasonic surgical aspirator system. It includes safety information, system overview, console and handpiece components, setup procedures, and troubleshooting.
The guide also contains technical specifications, sterilization validation, warranty details, and reference tables for handpiece frequencies, arm adjustments, and electrical requirements, helping users maintain safe and effective operation of the system.
| Handpiece frequency | 23 kHz and 36 kHz |
|---|---|
| Arm height positions | Normal, 7.6 cm above normal, 15.2 cm above normal |
| Mains plug requirement | 3‑pin plug with protective earth (or 2‑pin with integral earth) |
The system uses handpieces that operate at 23 kHz and 36 kHz.
The arm has three positions—normal, 7.6 cm above normal, and 15.2 cm above normal—adjusted by pressing a release tab and moving the arm.
Do not use the system in the presence of flammable anesthetics or any potentially explosive atmosphere.