About this manual

The Nebulizer Adjustment Fixture Kit (p/n G1960-67470) is covered by this user guide. The kit includes the Nebulizer Adjustment Fixture (GT430-20470), an LED Flashlight (2140-0826), a 40× Magnifier (1000-1541), and the user guide. It is used to adjust the Microflow Nebulizer Assembly (G1946-67260), ES Tested Nebulizer (G1958-67098), Tested ESI Nebulizer Assembly (G1946-67098), and Tested APCI Nebulizer Assembly (G1946-67037).

The guide explains how to insert and inspect a nebulizer, check needle centering, plugging, nozzle damage, needle length, and corrosion. It then gives procedures for adjusting or replacing the nebulizer needle, including tools and replacement kits, and provides troubleshooting guidance and recommended needle adjustment distances.

What's inside

  • Before you begin
  • To inspect the nebulizer
  • To adjust or replace the nebulizer needle
  • Tools needed
  • Parts needed
  • Needle replacement kits
  • Recommended needle adjustment distance
  • Troubleshooting

Specifications

Kit part numberG1960-67470
Nebulizer Adjustment Fixture part numberGT430-20470
LED Flashlight part number2140-0826
40× Magnifier part number1000-1541
Recommended needle adjustment distance, Microflow Nebulizer Assembly0.003 inch
Recommended needle adjustment distance, ES Tested Nebulizer0.000 (flush)
Recommended needle adjustment distance, Tested ESI Nebulizer Assembly0.003 inch
Recommended needle adjustment distance, Tested APCI Nebulizer Assembly0.000 (flush)

Frequently asked questions

Which nebulizers can be adjusted with the Nebulizer Adjustment Fixture Kit?

It is used with the Microflow Nebulizer Assembly (G1946-67260), ES Tested Nebulizer (G1958-67098), Tested ESI Nebulizer Assembly (G1946-67098), and Tested APCI Nebulizer Assembly (G1946-67037).

What are common symptoms of a plugged needle?

Common symptoms include increased LC back pressure, off-axis spraying, dripping from the nebulizer, and the reference nebulizer needle not spraying.

What is the recommended needle adjustment distance for an ES Tested Nebulizer?

The recommended distance is 0.000 inch (flush).

What should I do if the needle is clogged with salt?

Sonicate the nebulizer assembly in water and methanol, keeping the nozzle tip from touching the beaker or flask; replace the needle if it remains clogged.