This document describes the user guide for the nebulizer adjustment fixture.
The Nebulizer Adjustment Fixture Kit (p/n G1960-67470) is covered by this user guide. The kit includes the Nebulizer Adjustment Fixture (GT430-20470), an LED Flashlight (2140-0826), a 40× Magnifier (1000-1541), and the user guide. It is used to adjust the Microflow Nebulizer Assembly (G1946-67260), ES Tested Nebulizer (G1958-67098), Tested ESI Nebulizer Assembly (G1946-67098), and Tested APCI Nebulizer Assembly (G1946-67037).
The guide explains how to insert and inspect a nebulizer, check needle centering, plugging, nozzle damage, needle length, and corrosion. It then gives procedures for adjusting or replacing the nebulizer needle, including tools and replacement kits, and provides troubleshooting guidance and recommended needle adjustment distances.
| Kit part number | G1960-67470 |
|---|---|
| Nebulizer Adjustment Fixture part number | GT430-20470 |
| LED Flashlight part number | 2140-0826 |
| 40× Magnifier part number | 1000-1541 |
| Recommended needle adjustment distance, Microflow Nebulizer Assembly | 0.003 inch |
| Recommended needle adjustment distance, ES Tested Nebulizer | 0.000 (flush) |
| Recommended needle adjustment distance, Tested ESI Nebulizer Assembly | 0.003 inch |
| Recommended needle adjustment distance, Tested APCI Nebulizer Assembly | 0.000 (flush) |
It is used with the Microflow Nebulizer Assembly (G1946-67260), ES Tested Nebulizer (G1958-67098), Tested ESI Nebulizer Assembly (G1946-67098), and Tested APCI Nebulizer Assembly (G1946-67037).
Common symptoms include increased LC back pressure, off-axis spraying, dripping from the nebulizer, and the reference nebulizer needle not spraying.
The recommended distance is 0.000 inch (flush).
Sonicate the nebulizer assembly in water and methanol, keeping the nozzle tip from touching the beaker or flask; replace the needle if it remains clogged.