The Agilent Fragment Analyzer systems are automated capillary electrophoresis platforms for scalable, flexible, fast, and reliable electrophoresis of nucleic acids.
This Quick Guide describes the Agilent DNF-935 Reagent Kit, which is for the fast analysis of dsDNA fragments between 100 bp and 1,500 bp on the Agilent 5200, 5300, and 5400 Fragment Analyzer systems. Sizing and relative quantification between samples can be obtained using this kit, with example applications including general PCR fragment sizing and QC, and genotyping.
The guide lists the kit components for the 500-sample and 1000-sample kits, additional equipment and reagents required for analysis, essential measurement practices, sample plate preparation, gel/dye mixture volume tables for each system, and the dsDNA 935 assay operating procedure. It also includes analytical and physical specifications such as DNA sizing range, sizing accuracy, separation resolution, run times, and kit stability.
| DNA Sizing Range | 100 bp – 1,500 bp |
|---|---|
| DNA Sizing Accuracy | + 15% or better |
| DNA Fragment Concentration Range | 0.5 ng/µL – 50 ng/µL input DNA (adjustable by dilution sample) |
| Separation Resolution | 1 bp – 100 bp ≤ 15%; 100 bp – 1,500 bp ≤ 10% (22-47 and 33-55 Array); 1 bp – 1,500 bp ≤ 5% (55-80 Array) |
| DNA Sizing Precision | 2% CV |
| Total electrophoresis run time | 22 cm: 15 minutes, 33 cm: 20 minutes, 55 cm: 35 minutes |
| Samples per run | 12, 48 or 96; depending on the instrument type |
| Sample volume required | 2 µL (adjustable depending on sample concentration) |
| Kit stability | 4 months |
| Ambient operating temperature | 19 – 25 °C (66 – 77 °F) |
The Quick Guide is intended for use with the Agilent 5200, 5300, and 5400 Fragment Analyzer systems only.
The kit is for the fast analysis of dsDNA fragments between 100 bp and 1,500 bp.
Kit stability is 4 months.
Allow reagents to equilibrate at room temperature for 30 minutes prior to use.