This is a guide on how to use a tonometer. It covers how to measure intraocular pressure, and how to change the probe.
This quick guide covers loading a single-use probe, measuring intraocular pressure, changing settings for different species, and cleaning the probe base. It states that the probe tip should be 4 to 8 millimeters from the cornea and that six correct measurements are needed for a result.
The guide also describes button functions, sound signals for correct and incorrect measurements, automatic switch-off after two minutes of inactivity, and the steps for cleaning and reinserting the probe base.
| Probe-to-cornea distance | 4 to 8 millimeters (1/6 to 1/3 inch) |
|---|---|
| Measurements per result | Six correct measurements |
| Automatic switch-off | After 2 minutes of inactivity |
| Species settings | do = dog/cat; ho = horse |
Keep the distance from the tip of the probe to the patient’s cornea at 4 to 8 millimeters (1/6 to 1/3 inch).
Six correct measurements are needed before the IOP is shown on the display.
Unscrew the collar, remove the probe base, inject alcohol solution through it, dry it with clean air, then reinsert and screw the collar back.
One short beep indicates a correct single measurement, a double beep indicates an incorrect measurement, and a long beep indicates six successful measurements with an available result.