This user guide contains information on the preparation, operation and maintenance of the Thermo Scientific Orion ORP electrodes. These ORP electrodes combine a platinum redox sensing electrode and a silver/silver chloride reference electrode in one body. The redox/ORP electrodes can be used to make direct measurements of redox potentials or perform potentiometric titrations. The ORP Triode electrodes have a built-in thermistor for measuring temperature.
This user guide from Thermo Fisher Scientific covers the preparation, operation and maintenance of the Thermo Scientific Orion redox/ORP electrodes. These electrodes combine a platinum redox sensing electrode and a silver/silver chloride reference electrode in one body and can be used for direct measurements of redox potentials or potentiometric titrations.
The guide covers four electrode families: epoxy Sure-Flow combination electrodes (Cat. No. 9678BNWP and 967800) with a clog-resistant junction for dirty, viscous or wastewater samples; glass combination electrodes (Cat. No. 9778BNWP and 9778SC) for use with organic solvents; refillable ORP/ATC Triodes (Cat. No. 9180BNMD and 9180BN) with a built-in temperature probe; and low maintenance gel-filled ORP/ATC Triodes (Cat. No. 9179BNMD and 9179BN). It also lists required equipment, sample requirements, calibration with Orion ORP standard, measurement techniques and troubleshooting.
| Sensing element | Platinum redox sensing electrode |
|---|---|
| Reference electrode | Silver/silver chloride (Ag/AgCl) |
| Sure-Flow junction | Prevents clogging in dirty, viscous or wastewater samples (9678BNWP, 967800) |
| Body material | Epoxy (9678, 9180, 9179) or glass (9778) |
| Built-in temperature probe | Thermistor in ORP/ATC Triode electrodes (9180BNMD, 9180BN, 9179BNMD, 9179BN) |
| Reference solution | Gel-filled, no filling solution required (9179BNMD, 9179BN) |
| Meter connection | BNC or U.S. standard connection; adapter cable for other inputs |
Use Cat. No. 900011 filling solution for samples with a total ionic strength greater than 0.2 M, and Cat. No. 900001 for samples with a total ionic strength less than 0.2 M. No filling solution is required for the 9179BNMD or 9179BN low maintenance Triode.
No. Redox/ORP electrodes are not designed for constant current or amperometric titrations such as Karl Fischer moisture analysis.
Place the electrode in Orion ORP standard and wait for the reading to stabilize; the mV potential should be 220 mV ± 60 mV. If the reading is out of range, repeat with fresh standard, then clean or refill the electrode.
Use a glass body electrode, Cat. No. 9778BNWP or 9778SC, since epoxy body electrodes should only be used in aqueous solutions.