Instruction manual for Agilent 7667A Mini TD, introducing the features, installation and operation of the instrument
The Agilent 7667A Mini Thermal Desorber (Mini TD) is a single-stage thermal desorber. It has a basic G4370A model without a sampling pump for offline sampling, mainly used with the 7820A GC, and an enhanced G4370M model with a built-in sampling pump for online automatic gas sampling, mainly used with the 5975T LTM GC/MSD.
This installation and operation manual covers the instrument and its workflow, installation and operation, first startup, software operation, routine maintenance, and troubleshooting. It also describes control panel operation, method creation and saving, consumable replacement, and error messages. This manual is written in Chinese.
| Product type | Single-stage thermal desorber (Mini TD) |
|---|---|
| Models | G4370A (basic) and G4370M (enhanced) |
| G4370A sampling | Offline sampling; no sampling pump |
| G4370M sampling | Online automatic gas sampling; built-in sampling pump |
| Operating modes | Sampling, desorption, and aging |
| Carrier gas upstream pressure | Greater than 60 psi (or 0.4 MPa) |
| Maximum sample boiling point | Not above n-alkane C16 |
| Main GC compatibility for G4370A | 7820A GC |
| Main GC/MSD compatibility for G4370M | 5975T LTM GC/MSD |
It has no sampling pump, is used for offline sampling, and is mainly used with the 7820A GC.
It has a built-in sampling pump and supports online automatic gas sampling, mainly with the 5975T LTM GC/MSD.
It has three operating modes: sampling, desorption, and aging.
The carrier gas upstream pressure must be greater than 60 psi (or 0.4 MPa).