This manual introduces the features and characteristics of the PHA series high-voltage differential probe, usage methods, safety precautions, common faults and troubleshooting.
The RIGOL PHA series high-voltage differential probes (PHA0150, PHA1150, PHA2150) convert high differential voltages to low voltages for measurement with oscilloscopes or other waveform instruments. They feature bandwidths up to 200 MHz, two attenuation ratios (50X and 500X), a 5 MHz bandwidth limit, and audible/visual overload alarms.
The manual includes safety precautions, general inspection, probe overview, parts description, usage instructions, detailed specifications (technical parameters, dimensions, operating environment), and appendices with maximum differential voltage vs. frequency curves, package list, and warranty summary. This manual is written in Chinese.
| Bandwidth (-3 dB) | PHA0150: 70 MHz; PHA1150: 100 MHz; PHA2150 50X: 160 MHz, 500X: 200 MHz |
|---|---|
| Rise Time | PHA0150: ≤5 ns; PHA1150: ≤3.5 ns; PHA2150 50X: ≤2.4 ns (typ. ≤2.2 ns), 500X: ≤2 ns (typ. ≤1.8 ns) |
| Attenuation Ratios | 50X / 500X |
| Maximum Differential Measurement Voltage (DC + peak AC) | 50X: ±150 V; 500X: ±1500 V |
| Common Mode Voltage (DC + peak AC) | ±1500 V |
| Maximum Input to Ground Voltage (Vrms) | 600 V CAT III, 1000 V CAT II |
| Input Impedance | Single-ended to ground: 5 MΩ ±1%; Differential: 10 MΩ ±1% |
| Input Capacitance | Single-ended to ground: <4 pF; Differential: <2 pF |
| CMRR | DC >80 dB; 100 kHz >60 dB; 1 MHz >50 dB; 3.2 MHz >30 dB; 50 MHz >26 dB |
| Noise (Vrms) | 50X: <50 mV; 500X: <300 mV |
| Overload Indicator Voltage Threshold | 50X: 153 V ±3 V; 500X: 1530 V ±30 V |
| Delay Time | PHA0150: 21 ns; PHA1150: 16 ns |
| Bandwidth Limit (5 MHz) | ≥-3 dB @ 5 MHz |
| Overload Indicator (Red LED) | Yes |
| Audible Overrange Alarm | Yes (ON/Off selectable) |
| Auto Save Function | Yes |
| Output Impedance | 50 Ω |
| Safety Standards | EN61010-1:2010 |
| EMC Standards | EN61326-1:2013, EN61000-3-3:2013, EN61000-3-2:2006+A1:2009+A2:2009 |
| Probe Dimensions | 176 mm × 65 mm × 25 mm |
| Probe Body Weight | 216 g |
| High-Voltage Differential Input Cable Length | PHA0150/PHA1150: 28 cm; PHA2150: 17 cm |
| Probe Output Cable Length | 1 m |
| Operating Temperature | 0°C to 50°C |
| Storage Temperature | -30°C to 70°C |
| Operating Humidity | ≤85% RH |
| Storage Humidity | ≤90% RH |
| Operating Altitude | 3000 m |
| Storage Altitude | 12000 m |
The PHA series probes offer two attenuation ratios: 50X and 500X.
The PHA1150 has a bandwidth of 100 MHz.
The probe complies with EN61010-1:2010.