RIGOL PHA Series High-Voltage Differential Probe User Manual

This manual introduces the features and characteristics of the PHA series high-voltage differential probe, usage methods, safety precautions, common faults and troubleshooting.


RIGOL PHA Series High-Voltage Differential Probe User Manual - cover page
Brand
Rigol
Category
Probes
Document type
User Manual
Language
Chinese
Pages
40
File format
PDF
File size
1 MB
Published
31 October, 2023
Updated
31 October, 2023
MD5 checksum
914079DB9BA2E156936F99CFC9B09E88

About this manual

The RIGOL PHA series high-voltage differential probes (PHA0150, PHA1150, PHA2150) convert high differential voltages to low voltages for measurement with oscilloscopes or other waveform instruments. They feature bandwidths up to 200 MHz, two attenuation ratios (50X and 500X), a 5 MHz bandwidth limit, and audible/visual overload alarms.

The manual includes safety precautions, general inspection, probe overview, parts description, usage instructions, detailed specifications (technical parameters, dimensions, operating environment), and appendices with maximum differential voltage vs. frequency curves, package list, and warranty summary. This manual is written in Chinese.

What's inside

  • General Inspection
  • High-Voltage Differential Probe Overview
  • Probe Parts Overview
  • Parts Description
  • Probe Usage Method
  • Specifications
  • Appendix

Specifications

Bandwidth (-3 dB)PHA0150: 70 MHz; PHA1150: 100 MHz; PHA2150 50X: 160 MHz, 500X: 200 MHz
Rise TimePHA0150: ≤5 ns; PHA1150: ≤3.5 ns; PHA2150 50X: ≤2.4 ns (typ. ≤2.2 ns), 500X: ≤2 ns (typ. ≤1.8 ns)
Attenuation Ratios50X / 500X
Maximum Differential Measurement Voltage (DC + peak AC)50X: ±150 V; 500X: ±1500 V
Common Mode Voltage (DC + peak AC)±1500 V
Maximum Input to Ground Voltage (Vrms)600 V CAT III, 1000 V CAT II
Input ImpedanceSingle-ended to ground: 5 MΩ ±1%; Differential: 10 MΩ ±1%
Input CapacitanceSingle-ended to ground: <4 pF; Differential: <2 pF
CMRRDC >80 dB; 100 kHz >60 dB; 1 MHz >50 dB; 3.2 MHz >30 dB; 50 MHz >26 dB
Noise (Vrms)50X: <50 mV; 500X: <300 mV
Overload Indicator Voltage Threshold50X: 153 V ±3 V; 500X: 1530 V ±30 V
Delay TimePHA0150: 21 ns; PHA1150: 16 ns
Bandwidth Limit (5 MHz)≥-3 dB @ 5 MHz
Overload Indicator (Red LED)Yes
Audible Overrange AlarmYes (ON/Off selectable)
Auto Save FunctionYes
Output Impedance50 Ω
Safety StandardsEN61010-1:2010
EMC StandardsEN61326-1:2013, EN61000-3-3:2013, EN61000-3-2:2006+A1:2009+A2:2009
Probe Dimensions176 mm × 65 mm × 25 mm
Probe Body Weight216 g
High-Voltage Differential Input Cable LengthPHA0150/PHA1150: 28 cm; PHA2150: 17 cm
Probe Output Cable Length1 m
Operating Temperature0°C to 50°C
Storage Temperature-30°C to 70°C
Operating Humidity≤85% RH
Storage Humidity≤90% RH
Operating Altitude3000 m
Storage Altitude12000 m

Frequently asked questions

What are the attenuation ratios of the RIGOL PHA series probes?

The PHA series probes offer two attenuation ratios: 50X and 500X.

What is the bandwidth of the PHA1150 model?

The PHA1150 has a bandwidth of 100 MHz.

What safety standard does the probe comply with?

The probe complies with EN61010-1:2010.

Download PDF (1 MB)