This is a quick guide for using the Agilent 5200, 5300, and 5400 Fragment Analyzer systems to size and quantify small to mid-range molecular weight dsDNA smears/fragments.
This Quick Guide describes the Agilent DNF-493 HS Large Fragment Kit, intended for use with the Agilent 5200, 5300, and 5400 Fragment Analyzer systems. The kit is designed for the sizing and quantification of small to mid-range molecular weight dsDNA smears or fragments, with example applications including quality control of Next Generation Sequencing (NGS) libraries.
| DNA Sizing Range | 75 bp – 20,000 bp |
|---|---|
| DNA Sizing Accuracy | + 15% or better |
| DNA Fragment Concentration Range | 0.005 ng/µL – 0.5 ng/µL input DNA |
| DNA Smear Concentration Range | 0.05 ng/µL – 5 ng/µL input DNA |
| DNA Quantification Accuracy | + 25% |
| DNA Quantification Precision | 20% CV |
| Maximum DNA Concentration | 0.5 ng/µL per fragment; 5 ng/µL per total sample |
| Total electrophoresis run time (22cm) | 30 minutes |
| Total electrophoresis run time (33cm) | 50 minutes |
| Total electrophoresis run time (55cm) | 80 minutes |
| Kit stability | 4 months |
The kit is intended for use with the Agilent 5200, 5300, and 5400 Fragment Analyzer systems only.
The assay covers a DNA sizing range of 75 bp to 20,000 bp.
The maximum DNA concentration is 0.5 ng/µL per fragment.