This manual introduces the use of Agilent 5100 and 5110 ICP-OES and its functions
This manual provides comprehensive instructions for the Agilent 5100 and 5110 ICP-OES instruments. It begins with detailed safety practices addressing plasma hazards, heat, compressed gases, electrical risks, and required interlocks. The guide then covers site preparation requirements, instrument overview, and installation procedures including torch assembly, accessory connections, and software setup.
The operation section explains analysis checklists, instrument startup, method development, sample running, and reporting. Maintenance chapters detail routine cleaning, torch cleaning procedures, troubleshooting, spare parts, and technical support contacts. The document applies to both the 5100 and 5110 models unless otherwise specified.
| Plasma temperature | about 10,000 °C |
|---|---|
| Work coil voltage | 1,500 V RMS |
| Work coil frequency | about 27 MHz |
| Torch cool-down time | up to five minutes |
The plasma operates at about 10,000 °C.
The work coil operates at 1,500 V RMS and about 27 MHz.
Extinguish the plasma by pressing SHIFT + F5 or clicking the 'Plasma Off' icon in ICP Expert software before opening the door.
Allow up to five minutes for the torch and compartment to cool.