This manual provides information about the features, specifications, installation, operation, and maintenance of the Thermo Scientific Forma Direct Heat CO2 Incubator Model 310 Series.
The manual provides operating and maintenance instructions for the Forma Direct Heat CO2 Incubator Model 310 Series, which includes models with thermal conductivity (T/C) or infrared (IR) CO2 sensors, 115V or 230V power, and maximum temperatures of 45–55°C. It details intended use for tissue and cell culture, including assisted reproductive procedures, and outlines safety warnings for electrical hazards, asphyxiation risk from CO2/N2, and lifting requirements for the >200 lb unit.
The document covers installation and start-up, control panel operation, calibration procedures for temperature, CO2 (both sensor types), and relative humidity, configuration settings for alarms and communications, and alarm troubleshooting. It also includes warranty terms, service contacts, and a revision history noting updates to parts lists, schematics, and specifications such as temperature uniformity of 0.3°C.
| CO2 Sensor Type | Thermal conductivity (T/C) or Infrared (IR) |
|---|---|
| Voltage Options | 115V or 230V, 50/60 Hz |
| Maximum Temperature | 55°C (models 310/311), 50°C (320/321), 45°C (350/351/360/361) |
| Chamber Weight | >200 lbs (91 kg) |
| Temperature Uniformity | ±0.3°C |
| Intended Use | Incubation of tissue and cell samples; approved for microbiological incubator (21 CFR 866.2540) and assisted reproductive procedures (21 CFR 884.6120) |
The 310 Series offers either a thermal conductivity (T/C) sensor or an infrared (IR) sensor for CO2 control.
Models 310 and 311 reach 55°C; models 320 and 321 reach 50°C; models 350, 351, 360, and 361 reach 45°C.
The incubator weighs more than 200 lbs (91 kg) and a four-person lift is recommended.
No, calibration of control parameters is not covered under warranty; it is a billable service.