Dytran Model 5310M1 High Temperature Vibration Measurement System Supplemental Operating Instructions

This is an operating guide for Model 5310M1 high temperature vibration measurement system


Dytran Model 5310M1 High Temperature Vibration Measurement System Supplemental Operating Instructions - cover page
Brand
Dytran
Category
Industrial Robot
Document type
User Manual
Language
English
Pages
3
File format
PDF
File size
31 KB
Published
21 October, 2023
Updated
21 October, 2023
MD5 checksum
288E761486E504CEBA4F2171781ADDCB

About this manual

The Dytran Model 5310M1 is a high temperature vibration measurement system designed for use in Environmental Stress Screening (ESS) chambers and general environmental test chambers. The system combines a Model 3030C1 charge mode miniature accelerometer with a Model 4705M13 in-line charge amplifier and a low-noise treated cable with an environmental sealing boot. It allows users to enjoy the simplicity of Low Impedance Voltage Mode (IEPE) operation with the high temperature performance of a charge mode piezoelectric accelerometer.

What's inside

  • Introduction
  • Description
  • Installation
  • Sealing Boot Installation
  • Powering
  • Operating the System
  • Mixing System Components
  • Maintenance and Repair

Specifications

Suggested Current Level4 mA
Accelerometer Sensitivity0.4 pC/G
Charge Amplifier Sensitivity25 mV/pC
Nominal System Sensitivity10 mV/G
Power Unit Current Limit Range2 to 20 mA

Frequently asked questions

What is the Dytran Model 5310M1 system especially suited for?

It is especially suited for use in ESS (Environmental Stress Screening) chambers and for environmental test chambers in general.

Can the signal/power jack be connected to a standard DC power supply?

No, under no circumstances should it be connected to a source of DC power that does not have current limiting, such as a battery or a DC power supply, as this will destroy the circuitry.

Should components from different 5310M1 systems be mixed?

No, do not mix components from different systems as the system sensitivity will be in doubt because each system is calibrated as an individual set.