This manual provides steps for installing Micro Motion ELITE flow and density sensors.
This manual provides instructions for planning, mounting, wiring, and grounding Micro Motion ELITE Coriolis Flow and Density Sensors. It includes safety messages, hazard warnings, and references to related documentation such as the Configuration Manual and product data sheets.
| Maximum cable length (9-wire to MVD transmitter) | 1,000 ft (305 m) |
|---|---|
| Maximum cable length (4-wire without Ex-approval) | 1,000 ft (305 m) |
| Minimum ambient temperature for electronics | -40 °F (-40 °C) |
| Maximum ambient temperature for electronics | 140 °F (60 °C) |
| Preferred flow direction for liquids and slurries | Upward |
No, there are no pipe run requirements for Micro Motion sensors. Straight runs of pipe upstream or downstream are unnecessary.
The purge time for CMFS150 sensors is 6 minutes at a rate of 20 ft³/h.