Micro Motion® ELITE® flow meters are high-precision Coriolis flow meters that are often used in the oil and gas industry in conjunction with volume provers. These guidelines are designed to aid in troubleshooting and operation of provers that are in use together with Micro Motion Coriolis meters. Proving results may vary if unstable process conditions exist during proving.
This document serves as a reference guide for the operation and troubleshooting of volume provers when used in conjunction with Micro Motion ELITE high-precision Coriolis flow meters. It is particularly relevant for applications in the oil and gas industry where these meters are paired with volume provers to verify accuracy.
The text details best practices for prover size selection, emphasizing that pass times should not be less than 0.5 seconds. It outlines critical process conditions required for successful proving, such as maintaining stable flow rate, density, temperature, and pressure. Additionally, it provides specific configuration settings for the meter transmitters to optimize filtering and response speed during proving runs.
The prover size should never result in a pass time of less than 0.5 seconds.
The minimum back pressure must be sufficient to maintain stable flow rate and avoid vapor breakout, calculated using Equation 4 from API MPMS Ch. 5.6.
Flow damping should be set to a value between 0.0 and 0.08 seconds.