This document is the user manual for RIGOL RP1000D series high-voltage differential probes, which introduces the warranty and declaration, trademark information, product certification, general safety overview and other contents of the product.
The RIGOL RP1000D series high-voltage differential probe converts high input differential voltages into low voltage signals for display on an oscilloscope. The series includes models with bandwidths up to 25MHz, 50MHz, and 100MHz, suitable for large power testing and R&D applications. This manual is written in Chinese.
| Bandwidth (RP1025D) | DC - 25MHz (-3dB) for X50/X200 ratios; DC - 15MHz for X20 ratio |
|---|---|
| Bandwidth (RP1050D) | DC - 50MHz (-3dB) for X200/X500/X1000 ratios; DC - 25MHz for X100 ratio |
| Bandwidth (RP1100D) | DC - 100MHz (-3dB) for X200/X500/X1000 ratios; DC - 25MHz for X100 ratio |
| Accuracy | ±2% |
| Input Impedance (Differential) | 4MΩ/1.2pF for RP1025D; 100MΩ/1.2pF for RP1050D/RP1100D |
| Output Impedance | 50Ω |
| Power Supply | Specified external 9V DC power supply |
The maximum differential voltage is 1400V (DC+AC peak-to-peak) or 450Vrms.
The maximum voltage to ground at the input terminal is 6500Vrms.
The kit includes a user guide, AC power adaptor, coaxial cable, dual-banana plug silicon cable, high voltage IC clip, safety alligator clip, and safety contact probe prod.