This note explores what parameters settings to use to obtain the best data and avoid signal loss when the LC/MSD is used in positive/negative switching mode.
This document addresses challenges in acquiring data from narrow chromatographic peaks (0.1-0.2 minutes) on the Agilent 1100 Series LC/MSD system, specifically regarding positive and negative ionization mode switching.
The text details experimental results showing that default acquisition parameters cause significant signal loss when using wide scan ranges (e.g., m/z 150-1150). It recommends adjusting the chromatographic peak width parameter from 0.1 to 0.3 minutes and disabling the time filter to recover signal abundance while maintaining data fidelity.
Increase the chromatographic peakwidth parameter from 0.1 to 0.3 minutes while disabling the time filter.
At least six to nine data points or scans need to be acquired across the peak.