Optimizing LC/MSD Acquisition Parameters for Positive/Negative Mode Switching Technical Note

This note explores what parameters settings to use to obtain the best data and avoid signal loss when the LC/MSD is used in positive/negative switching mode.


Optimizing LC/MSD Acquisition Parameters for Positive/Negative Mode Switching Technical Note - cover page
Brand
Agilent
Document type
Other
Language
English
Pages
5
File format
PDF
File size
69 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
14C77E14D8559E5A48ADC5C045FB0CE3

About this manual

This document addresses challenges in acquiring data from narrow chromatographic peaks (0.1-0.2 minutes) on the Agilent 1100 Series LC/MSD system, specifically regarding positive and negative ionization mode switching.

The text details experimental results showing that default acquisition parameters cause significant signal loss when using wide scan ranges (e.g., m/z 150-1150). It recommends adjusting the chromatographic peak width parameter from 0.1 to 0.3 minutes and disabling the time filter to recover signal abundance while maintaining data fidelity.

What's inside

  • Introduction
  • Experimental Setup
  • Results and Discussion
  • Chromatographic Conditions
  • MS Conditions
  • Table 1: Variation in Response
  • Conclusions

Frequently asked questions

What is the recommended adjustment for wide scan ranges to avoid signal loss?

Increase the chromatographic peakwidth parameter from 0.1 to 0.3 minutes while disabling the time filter.

How many scans are needed across a peak for chromatographic fidelity?

At least six to nine data points or scans need to be acquired across the peak.