Step-by-step upgrade of Agilent 1100 to 1200 Series Rapid Resolution LC systems Technical Note

This document introduces how to upgrade Agilent 1100 Series LC systems to Agilent 1200 Series RRLC systems for higher performance. The upgrade can be done through a step-by-step upgrade of Agilent 1100 Series system to Agilent 1200 Series RRLC system that is fully configured to provide: Shorter run and cycle times for increased productivity; More resolution for more certainty of not missing impurities; Improved signal-to-noise ratios for lower detection limits.


Step-by-step upgrade of Agilent 1100 to 1200 Series Rapid Resolution LC systems Technical Note - cover page
Brand
Agilent
Document type
Other
Language
English
Pages
8
File format
PDF
File size
413 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
8B509FB19354C0C8804A6AE5EBFE09DC

About this manual

Agilent Technologies presents a method to optimize existing Agilent 1100 Series LC systems by upgrading them step-by-step to the newer Agilent 1200 Series Rapid Resolution Liquid Chromatography (RRLC) technology. The document outlines four specific steps involving the replacement of columns, detectors, and column compartments to achieve faster results and better resolution.

The upgrade process begins with replacing a standard column packed with 5 µm particles with a shorter column packed with 1.8 µm particles. Subsequent steps involve swapping the detector for an Agilent 1200 Series diode array detector (DAD SL) and integrating a thermostatted column compartment designed to operate up to 100 °C. The final configuration includes a low-delay binary pump, resulting in significantly improved signal-to-noise ratios and resolution.

What's inside

  • Introduction
  • Step 1: Replacing the column
  • Step 2: Replacing the detector
  • Step 3: Replacing the column compartment
  • Final configuration

Frequently asked questions

What is the maximum operating temperature for the new column compartment?

The Agilent 1200 Series thermostatted column compartment SL is designed to operate up to 100 °C.

What was the peak-to-peak noise of the Agilent 1200 Series DAD SL?

The peak-to-peak noise for this application was 0.2928 mAU.

What is the internal volume of the new heat exchanger installed in step 3?

The new heat exchanger had an internal volume of 1.6 pL.