Agilent ELS UV Detection of Succinic Acid Technical Overview

Describes the advantages of Agilent Evaporative Light Scattering Detector (ELSD) over UV detector in analyzing succinic acid, including stable baselines, no drift and solvent peaks, and much greater response.


Agilent ELS UV Detection of Succinic Acid Technical Overview - cover page
Brand
Agilent
Category
Other
Document type
Product Leaflet
Language
English
Pages
3
File format
PDF
File size
584 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
30A02A431E81194265AFB4137533AAA1

About this manual

This document provides a technical overview of using the Agilent Evaporative Light Scattering Detector (ELS) for the analysis of succinic acid. It explains that succinic acid lacks a good UV chromophore, making UV detection difficult without denaturing the solute or requiring gradient elution which complicates separation from other substances like proteins.

The Agilent ELS detector is presented as a superior alternative, offering clean baselines, no drift, and no solvent peaks. The document details specific experimental conditions used to demonstrate these capabilities, including sample concentration, column specifications, eluent composition, and flow rate, showing that the ELS can detect succinic acid down to an on-column load of 2 yg.

What's inside

  • Introduction
  • Succinic Acid Structure
  • Detection Challenges
  • Experimental Conditions
  • UV vs ELS Comparison
  • Limits of Detection

Specifications

Sample Concentration1 mg/mL
Column SpecificationPLRP- 300A 8 ym, 4.6 x 150 mm
Eluent A Composition1% ACN, 0.1% TFA, 99% Water
Eluent B Composition100% ACN, 0.1% TFA
Gradient Profile5-80% B in 20 min
Flow Rate1.0 mL/min
UV Detection Wavelength215 nm
Minimum Detection Limit2 yg on-column load

Frequently asked questions

Why is UV detection difficult for succinic acid?

Succinic acid does not have a good UV chromophore, and separating it from substances requiring gradient elution rules out RI detection.

What are the advantages of using Agilent ELS over UV detection?

The Agilent ELS delivers clean baselines, no drift, no solvent peaks, and much greater response than a UV detector.

What is the minimum detection limit for succinic acid with Agilent ELSD?

It is possible to detect succinic acid down to an on-column load of 2 yg.